SYSTEM
CABINETS
(FOOTPRINT)
FRONT
NOTE 3
BACKBOARD
TERMINATION
FIELD
(NOTE 4)
TABLE
NOTES:
1. 115V AC, 60 Hz, 15 AMP OUTLETS
(HUBBELL 5262 OR EQUIVALENT)
MUST BE LOCATED WITHIN SIX FEET
(1.8
m) OF SYSTEM CABINETS.
2. MULTIPLE CABINET SYSTEMS REQUIRE
TWO QUAD OUTLETS, SINGLE CABINET
SYSTEMS REQUIRE ONE QUAD OUTLET.
3. ALLOW AT LEAST 24 INCHES OF SPACE
IN FRONT OF CABINETS. TABLE MUST
BE ABLE TO SUPPORT 250 POUNDS.
4. BACKBOARD IS 3/4 INCHES THICK BY
48 INCHES WIDE BY 72 INCHES LONG
Figure 6-1. Typical System 25 Equipment Area Floor Plan