Anritsu MT8852A Bluetooth Headset User Manual


 
MT8850A / MT8852A Bluetooth Test Set Using the Script Menu
10. Choose “Toggle Payload” and select either “Continuous” or “Once” as detailed below.
Continuous: This is the default setting in which, to comply with the RF test
specification, the payload is changed after each packet measured.
Once: This setting reduces the time required to complete the test. The
payload is changed only once per measurement stage. The payload
is changed after the selected number of packets has been measured
with the first payload and the selected number of packets is then
measured with the second payload.
11. Press the
[1 of 2] soft key to move to the second screen.
The second screen will vary according to whether the Test type is in Loopback or TX
mode. The following screen is in Loopback mode.
The following screen is in TX mode.
You can select or deselect the test to run and make changes to the frequency values
shown.
12. To select or deselect a test, choose the test and press the
key.
13. To change a frequency, choose the frequency, press the
key, enter the value
using the keypad and press the
[Enter] soft key.
14. Press the
[Defaults] soft key to restore the default values.
15. Press the
[Limits] soft key to display the Modulation index test limits screen.
16. Press the
[Defaults] soft key to restore the default values.
17. Choose the limit you want to change.
8-30 13000-000109
Testing in Standard Script Mode June 2003