Clarity - INT9 A/D converter Troubleshooting
29
• In the Models list select the DataApex
Acquisition device, Model 9.00 and press the
Next button.
• In the remaining dialogs press the Next
button.
• Restart the PC.
• Re-check the driver in the
Device Manager as
desribed in chapter 4.3.3.
4.4 Data Acquisition – non-functional
A gray icon, label “Disabled”, and Data Acquisition command not functional
Other manifestations of this error are also:
Method Setup - Acquisition tab missing, Method –
Acquisition command non-functional, Run,
Stop, Abort commands non-functional in the
windows
Single Analysis and Sequence).
Possible causes are:
a) Detector not allocated to instrument:
Open the System Configuration dialog from the
main
Clarity window using the System -
Configuration command and check the tab of
the corresponding instrument -
Instrument x. If it
has no allocated detectors, add them.
Solution: In the left-hand list Setup Control Modules
select the correct detector that is connected to
the A/D card you are using and link it to the
corresponding instrument on the right.