Agilent Technologies 5071A Wireless Office Headset User Manual


 
Operating and Programming
In Case of Difficulty
1-8
In Case of Difficulty
This section describes the additions, changes, and deletions to the
information that appears in chapter 3 of the 5071A Operating and
Programming manual.
Front-Panel Indicators and Messages
Table 3-1. Front-Panel Status Lights
All references to internal-battery power, battery status, or battery
operation are deleted.
Operating System Failure
Restarting the 5071A
Step 3 of the procedure is deleted.
Step 4 is renumbered and changed:
3. Reapply external dc or ac-line power.
Step 5 becomes step 4.
Verifying Operation
2. Rear-Panel Output Connectors
Checking the Outputs
Step 1 of the procedure is changed:
1. Connect one of the 5071A outputs to an input channel of the
54600A with a 50 feed through termination and BNC cable.
(When verifying rear-panel telecommunications outputs, use
manual scope settings and the appropriate termination as shown
on the rear-panel labels and in Table 1-1. The referenced rear-
panel telecommunications outputs should comply with ITU-T
(formerly CCITT) G.703/6 (G.703/2 if option 104 or 105 is present),
formatted per G.704. Sync Out complies per G.703/10, 75
(Additionally @ 1.544 MHz if option 104 or 105 is present.
Option 048:
Option 048:
Option 048:
Option 048:
Any TCO Option: