PROBLEM PROBABLE CAUSE CORRECTIVE MEASURES
4) Unusually high body fat readings
(40-50%)
a) Too much light present during the
reading.
a) Do not lift the Light Wand from the
subject's biceps until a "2" appears
on the screen during the first
measurement or the printer starts
printing after the second
measurement.
b) Use the Light Shield to block the
external light. Always avoid
measurements under strong lights
(e.g. direct sunlight) because errors
may occur.
5) Printer Stops a) Low batteries.
b) Bad battery contact.
c) Bad AC connection.
a) Replace batteries (see page 3).
b) Remove batteries and clean tabs
with isopropyl alcohol (rubbing
alcohol).
c) Unplug the AC Adapter from the
wall socket and from the unit. Clean
the adapter contact with isopropyl
alcohol (rubbing alcohol).