NOTE:
IT IS IMPORTANT TO: Aim the silver stripe of the Light Wand directly toward the shoulder. Keep the
arm relaxed (do not permit the subject to flex his/her muscles). A relaxed
biceps can be achieved by having the subject seated while resting his arm
(elbow bent) on a table.
Use the Light Shield to block the external light. Always avoid measurements
under strong lights (e.g. direct sunlight) because errors may occur.
CAUTION: Do not move the Light Wand while the decimal point is illuminated. The actual
measurement is being made at that time.
NOTE: Aim silver stripe directly toward the shoulder. Keep arm relaxed.
4f)
Press ENTER.
The display will flash a decimal point following the number "1a".
Then
the number "2a" will be displayed.
4g) Remove the Light Wand and replace it on the same point on the biceps. Press ENTER. The display will flash a
decimal point, then the number "1b".
4h) Place Light Wand firmly on Triceps of same arm (see page 15 for exact measurement point).
Press ENTER. The display will flash a decimal point. Don't move Light Wand until "2b" is displayed.
4i) Remove Light Wand and replace it on the same point on the triceps. Press ENTER. The display will flash a
decimal point. Don't move Light Wand until "1c" is displayed.