MODEL 6081-P pH/ORP SECTION 8.0
PROGRAMMING THE TRANSMITTER
32
8.9 CHOOSING A DISPLAY TIMEOUT
8.9.1 Purpose
The local transmitter screen will timeout to preserve power module life. The transmitter will continue to receive
measurement inputs and continue to transmit to the wireless network as programmed.
8.9.2 Procedure: Programming the display timeout
1. Press MENU to activate the live display.
2. Press MENU again. The menu screen appears. Choose Display.
3. Choose >>.
4. Select Display Timeout. Press ENTER.
5. Using the up and down keys, enter a value from 01 sec. to 999 sec. Press ENTER.
6. To return to the main display, press MENU then EXIT.