Emerson 6081-P Wireless Office Headset User Manual


 
MODEL 6081-P pH/ORP SECTION 2.0
INSTALLATION
SECTION 2.0
INSTALLATION
2.1 Considerations
2.2 Unpacking and Inspection
2.3 Pre-Installation Set Up
2.4 Mechanical Installation
2.5 Ground the Transmitter
2.6 Power Module Installation
2.1 CONSIDERATIONS
The transmitter can be commissioned before or after installation. It may be useful to commission it on the bench, before
installation, to ensure proper operation and to become familiar with its functionality. When applicable, make sure the instru-
ments are installed in accordance with intrinsically safe or non-incendive field wiring practices. The device will be powered
whenever the power module is installed. To avoid depleting the power module, make sure it is removed when the device
is not in use.
Power module
The Model 6081 is battery powered. The power module with the wireless unit contains 2 “C” size primary
lithium/thionyl chloride batteries. Each power module contains approximately .5 grams of lithium. Under normal
conditions, the power module materials are self-contained and are not reactive as long as the power module
integrity is maintained. Care should be taken to prevent thermal, electrical or mechanical damage. Contacts
should be protected to prevent premature discharge. Use caution when handling the power module. The power
module may be damaged if dropped from heights in excess of 20 feet.
Sensor
Make sensor connections through the cable entry in the enclosure. Be sure to provide adequate clearance
for cover removal.
Environmental
Verify that the operating atmosphere of the transmitter is consistent with the appropriate hazardous locations
certifications.
2.2 UNPACKING AND INSPECTION
Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If
there is no apparent damage, remove the transmitter. Be sure all items shown on the packing list are present. If
items are missing, immediately notify Rosemount Analytical.
Save the shipping container and packaging. They can be reused if it is later necessary to return the transmitter to the
factory.
5