19
MODEL 6081-P pH/ORP SECTION 6.0
DISPLAY AND OPERATION
SECTION 6.0
DISPLAY AND OPERATION
6.1. DISPLAY
The Model 6081 has a two-line
display. Generally, the user can pro-
gram the transmitter to show one of
three displays. If the transmitter has
been configured to measure ORP or
Redox, similar displays are avail-
able. Figure 6-1 shows the displays
available for pH.
The transmitter has information
screens that supplement the data in
the main display. Press to view
the information screens. The first
information screen shows the type
of measurement being made (pH,
ORP, Redox). The last information
screen is the software version
number.
During calibration and program-
ming, key presses cause different
displays to appear. The displays are
self-explanatory and guide the user
step-by-step through the procedure.
FIGURE 6-1. Displays During Normal Operation
Screen A shows the pH reading, the temperature, and the output current gen-
erated by the transmitter. Screen B shows the same information as Screen A
except the output current has been substituted with the raw sensor voltage.
Screen C is most useful while troubleshooting sensor problems.
6.1 Display
6.2 Keypad
6.3 Menus - pH
6.4 Information Screen Messages
6.5 Security
Screen A
Screen B
Screen C
Temperature
Temperature
Temperature
Output current
Input current
Glass
Impedance
Reference
Impedance
pH
pH
pH