8.5.2 Procedure: Setting a security code.
1. Press MENU. The menu screen appears. Choose Program.
2. Choose >>.
3. Choose Security.
4. Choose Calib or Config.
a. If you chose Calib, enter a three-digit security code.
b. If you chose Config, enter a three-digit security code.
5. To return to the main display, press MENU the EXIT.
8.5 SETTING A SECURITY CODE
8.5.1 Purpose
This section describes how to set a security code. There are three levels of security:
a. A user can view the default display and information screens only.
b. A user has access to the calibration menus only.
c. A user has access to all menus.
The security code is a three-digit number. The table shows what happens when security codes are assigned to
Calib (calibration) and Config (configure). In the table XXX and YYY are the assigned security codes.
Code assignments
Calib Config What happens
000 XXX User enters XXX and has access to all menus.
XXX YYY User enters XXX and has access to calibration menus only. User enters YYY and has access to
all menus.
XXX 000 User needs no security code to have access to all menus.
000 000 User needs no security code to have access to all menus.