Maintaining the Equipment
October 2002 4-5
CAUTION: To avoid having chemicals overflow from the effluent bottles, be
sure that the volume of chemicals in the replenishment bottles
does not exceed the remaining capacity of the effluent bottles.
Cleaning the Crossover Assemblies
Clean the crossover assemblies in the processor with a soft cloth to remove any
chemical residue on them. Install the splash guard see “Preventing Algae (for
Wash Configuration)” on page 4-8. Rinse the gears and socks with warm water
from a squirt bottle to dissolve deposits.
Operate the drive for the processor for at least 30 minutes before making prints or
running a control strip. This will ensure that the socks are thoroughly soaked with
chemistry.