Fluke 3458/HFL Wireless Office Headset User Manual


 
AC Voltage ±(ppm of setting + ppm of range), relative to calibration standards used.
Calibration Interval 365 day, ±1°C of last TCAL
Range Frequency 1281 HP3458/HFL 8508A
12.00000mV 1Hz - 40Hz - 300 + 300 -
40Hz - 1kHz - 200 + 110 -
1kHz - 20kHz - 300 + 110 -
20kHz - 50kHz - 1000 + 110 -
50kHz - 100kHz - 5000 + 110 -
100kHz - 300kHz - 40000 + 200 -
120.0000mV 1Hz - 40Hz - 70 + 40 -
40Hz - 1kHz - 70 + 20 -
1kHz - 20kHz - 140 + 20 -
20kHz - 50kHz - 300 + 20 -
50kHz - 100kHz - 800 + 20 -
100kHz - 300kHz - 3000 + 100 -
300kHz - 1MHz - 10000 + 100 -
1MHz - 2MHz - 15000 + 100 -
199.9999mV 1Hz - 10Hz - - 120 + 70
10Hz - 40Hz - - 120 + 70
40Hz - 100Hz 100 + 20 - 100 + 20
100Hz - 2kHz 100 + 20 - 100 + 10
2kHz - 10kHz 100 + 20 - 100 + 20
10kHz - 30kHz 300 + 40 - 300 + 40
30kHz - 100kHz 700 + 100 - 700 + 100
1.2000000V 1Hz - 40Hz - 70 + 40 -
40Hz - 1kHz - 70 + 20 -
1kHz - 20kHz - 140 + 20 -
20kHz - 50kHz - 300 + 20 -
50kHz - 100kHz - 800 + 20 -
100kHz - 300kHz - 3000 + 100 -
300kHz - 1MHz - 10000 + 100 -
1MHz - 2MHz - 15000 + 100 -
1.999999V 1Hz - 10Hz - - 100 + 60
10Hz - 40Hz - - 100 + 10
40Hz - 100Hz 80 + 10 - 80 + 10
100Hz - 2kHz 60 + 10 - 60 + 10
2kHz - 10kHz 80 + 10 - 80 + 10
10kHz - 30kHz 200 + 20 - 200 + 20
30kHz - 100kHz 500 + 100 - 500 + 100
100kHz - 300kHz 3000 + 1000 - 3000 + 1000
300kHz - 1MHz 10000 + 10000 - 10000 + 10000
12.00000V 1Hz - 40Hz - 70 + 40 -
40Hz - 1kHz - 70 + 20 -
1kHz - 20kHz - 140 + 20 -
20kHz - 50kHz - 300 + 20 -
50kHz - 100kHz - 800 + 20 -
100kHz - 300kHz - 3000 + 100 -
300kHz - 1MHz - 10000 + 100 -
1MHz - 2MHz - 15000 + 100 -