Fluke 3458/HFL Wireless Office Headset User Manual


 
AC Current ±(ppm of setting + ppm of range), relative to calibration standards used.
Calibration Interval 365 day, ±1°C of last TCAL
Range Frequency 1281 HP3458/HFL 8508A
120.0000µA 10Hz - 20Hz - 4000 + 200 -
20Hz - 45Hz - 1500 + 200 -
45Hz - 100Hz - 600 + 200 -
100Hz - 5kHz - 600 + 200 -
199.9999µA 1Hz - 10Hz - - 250 + 100
10Hz - 5kHz 200 + 200 - -
10Hz - 10kHz - - 250 + 100
10kHz - 30kHz - - 600 + 100
30kHz - 100kHz - - 4000 + 100
1.2000000mA 10Hz - 20Hz - 4000 + 200 -
20Hz - 45Hz - 1500 + 200 -
45Hz - 100Hz - 600 + 200 -
100Hz - 5kHz - 300 + 200 -
5kHz - 20kHz - 600 + 200 -
20kHz - 50kHz - 4000 + 400 -
50kHz - 100kHz - 5500 + 1500 -
1.999999mA 1Hz - 10Hz - - 250 + 100
10Hz - 5kHz 200 + 100 - -
10Hz - 10kHz - - 250 + 100
10kHz - 30kHz - - 600 + 100
30kHz - 100kHz - - 4000 + 100
12.00000mA 10Hz - 20Hz - 4000 + 200 -
20Hz - 45Hz - 1500 + 200 -
45Hz - 100Hz - 600 + 200 -
100Hz - 5kHz - 300 + 200 -
5kHz - 20kHz - 600 + 200 -
20kHz - 50kHz - 4000 + 400 -
50kHz - 100kHz - 5500 + 1500 -
19.99999mA 1Hz - 10Hz - - 250 + 100
10Hz - 5kHz 200 + 100 - -
10Hz - 10kHz - - 250 + 100
10kHz - 30kHz - - 600 + 100
30kHz - 100kHz - - 4000 + 100
120.0000mA 10Hz - 20Hz - 4000 + 200 -
20Hz - 45Hz - 1500 + 200 -
45Hz - 100Hz - 600 + 200 -
100Hz - 5kHz - 300 + 200 -
5kHz - 20kHz - 600 + 200 -
20kHz - 50kHz - 4000 + 400 -
50kHz - 100kHz - 5500 + 1500 -
199.9999mA 1Hz - 10Hz - - 250 + 100
10Hz - 5kHz 200 + 100 - -
10Hz - 10kHz - - 250 + 100
10kHz - 30kHz - - 600 + 100
1.050000A 10Hz - 20Hz - 4000 + 200 -
20Hz - 45Hz - 1600 + 200 -
45Hz - 100Hz - 800 + 200 -
100Hz - 5kHz - 1000 + 200 -
5kHz - 20kHz - 3000 + 200 -
20kHz - 50kHz - 10000 + 400 -
1.999999A 10Hz - 1kHz 500 + 200 - -
10Hz - 2kHz - - 600 + 100
1kHz - 5kHz 1500 + 400 - -
2kHz - 10kHz - - 700 + 100
10kHz - 30kHz - - 3000 + 100
19.99999A 10Hz - 2kHz - - 800 + 100
2kHz - 10kHz - - 2500 + 100