Fluke 3458/HFL Wireless Office Headset User Manual


 
Resistance
±(ppm of setting + ppm of range), relative to calibration standards used.
Calibration Interval 365 day, ±1°C of last TCAL
Range 1281 HP3458/HFL 8508A
1.99999999 Ohm - - 10 + 2.0
12.00000 Ohm - 10 + 5 -
19.9999999 Ohm 12 + 1.0 - 7.0 + 0.7
120.00000 Ohm - 8 + 5 -
199.999999 Ohm 8 + 0.3 7.0 + 0.25
1.2000000 kOhm - 7.5 + 0.5 -
1.99999999 kOhm 6 + 0.3 - 7.0 + 0.25
12.000000 kOhm - 7.5 + 0.5 -
19.9999999 kOhm 6 + 0.3 - 7.0 + 0.25
120.00000 kOhm - 7.5 + 0.5 -
199.999999 kOhm 6 + 0.3 - 7.0 + 0.25
1.2000000 MOhm - 15 + 2 -
1.99999999 MOhm 10 + 0.7 - 7.0 + 0.5
12.000000 MOhm - 50 + 10 -
19.9999999 MOhm 20 + 4.0 - 9.0 + 5.0
120.00000 MOhm - 500 + 10 -
199.999999 MOhm 200 + 45 - 30 + 50
1.2000000 GOhm - 5000 + 10 -
1.99999999 GOhm 2000 + 450 - 500 + 500